When operating the system, please follow these steps:

1- Open the program icon on the desktop to show the program menu as shown in the picture (A).
2- Pressing the “Port setting” icon from the list as shown in the image (A) to show the screen as shown in the image (B).
3- Press No. (1) (Scan Ports) for (Compressor Boards) in the picture (B) and select the violator (Com1).
4- Press the number (2) (Scan Ports) for (Message Center) in the picture (B) and choose the opposite of (Com 1 & Com compressor boards).
5- Click on the Save icon and then click on Back.
6- The Run System icon in the image (A) is pressed to launch the main
screen.
7- After running the program, the status of the compressor and evaporator
and the temperature of each refrigerator or freezer separately will appear in red as shown in the picture (C).
8- The temperatures are set as follows: – The Setting icon is selected as in the picture (D) and then press No. (1) as shown in the picture (E) so that
the screen appears in the example as the picture (F).
9- The maximum and minimum temperature is set for each refrigerator
separately, as shown in the picture (F).
10- The data of the specialists working on this program is recorded by choosing the SMS setting icon to show the screen as shown in the picture (G) and recording the phone numbers of the specialists to receive text messages on their phones indicating if a refrigerator malfunction occurred or not.
11- To request the recall of the stored data and the printing of reports on the status of the compressors and the temperatures of the refrigerators, the Logger Data file icon is opened, which is installed next to the program on the desktop.
12- The file as shown in the picture (H) is divided into two parts, a section for the status of the staplers, with reports specified by date, and another
section with temperatures and also specified by dates.
13- To print a report for any of the staplers or temperatures, the report is
opened from the file as shown in the picture (J), then click on the file of the report as shown in the picture (I) and then press on Print to print the report registered in the system.

Note: Please choose the communication system with a good signal in the exact location to operate the system

As for the integrated monitoring system, the user can set the freezer temperature alarm at the desired temperature for the user less than the minimum defrost temperature by 5 degrees Celsius. If the freezer exceeds the permissible limit for temperature rise that occurs in the defrost stage in a period exceeding half an hour If there is a problem in the freezing circuit. Therefore, we recommend that the temperature be a warning of the high temperature (-15) of the freezer and send text messages to the phone numbers indicated by the user.

Determine the time point of the data
Data for one day = 2.093kb * 30 = 62.79kb * 12 = 753.43kb * 60 = 45208kb
As the storage space for Data Roger after deducting the program space is 700 GB.

Note: Please choose the communication system with a good signal in the exact location to operate the system